User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 5340A:Frequency Counter(CP14-026) DATE: 19-Sep-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 36 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 DISP Allow UUT a 1 hour warm-up. 1.004 DISP Connect the Mieco 1402 REFRENCE OUTPUT to the 8673C 1.004 DISP FREQ STANDARD EXT input and set the 8673C INT/EXT 1.004 DISP switch to the EXT position. 1.005 DISP Connect the 8673C 10 MHz OUT to the 3325A EXT REF IN 1.005 DISP connector. 1.005 DISP Set the UUT front panel controls as follows: 1.005 DISP RESOLUTION ..................... 1Hz 1.005 DISP SAMPLE RATE .................... fully ccw 1.005 DISP RANGE switch ................... 10Hz-250MHz 1.005 DISP INT-EXT switch ................. INT (rear panel) 1.006 CALL HP 5340A Sub1 1.007 CALL HP 5340A Sub2 1.008 CALL HP 5340A Sub3 1.009 HEAD 1.010 DISP Set the UUT rear panel controls as follows: 1.010 DISP INT-EXT switch ................. INT 1.011 HEAD {} 1.012 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.012 DISP {º THIS COMPLETES THE VERIFICATION º} 1.012 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.013 END